Paper
TENTH VALUE LAYERS FOR 60Co GAMMA RAYS AND FOR 4,
6, 10, 15, AND 18 MV X RAYS IN CONCRETE FOR BEAMS OF
CONE ANGLES BETWEEN 0° AND 14° CALCULATED BY
MONTE CARLO SIMULATION
Adnan K. Jaradat* and Peter J. Biggs†
shielding barriers using the methodology outlined in
National Council on Radiation Protection and Measurements (NCRP) Report No. 151 (NCRP 2005). This report
supercedes previous reports on this topic (NCRP 1976,
1977, 1984). This report contains the first and subsequent
tenth value layer (TVL) values for all energies used in
current therapy electron linear accelerators. However, the
published data are based on broad beam attenuation; that
is, when the jaws of the linear accelerator or 60Co unit are
fully open. For a linear accelerator, the typical maximum
field size is 40 ⫻ 40 cm2 at 100 cm and for 60Co it is 33 ⫻
33 cm2 at 80 cm.
Broad beam attenuation represents the most conservative shielding assumption because fields up to the
maximum size can be and are used. However, analysis of
field sizes used in the radiotherapy clinic (Biggs 1991b)
indicate that typical clinical field sizes are much less than
the maximum field size available from medical linear
accelerators. This analysis shows that 50% of field sizes
treated are less than 15 ⫻ 15 cm2 and 90% are less than
26 ⫻ 26 cm2. About the only disease site that routinely
requires large fields is Hodgkins’ disease where the chest
and lower neck is irradiated. However, even the effective
field sizes are much reduced by the lung blocks and this
patient population, mainly adolescents, is comparatively
small. These results clearly depend on a specific clinic
and the mix of patients treated, for example, prostate,
head and neck, breast, etc., but, on general grounds, it is
unlikely that results from a different clinic would be
dramatically higher. There has also been a trend in
radiation therapy in specific disease sites to reduce the
treatment field as much as possible. One of the reasons
for this field reduction, made possible, in part, by better
imaging for localization in recent years, is to reduce the
co-morbidity of the treatment and the possibility of
second cancers, given that many patients, in particular
pediatric patients, can live for extended periods. For
example, in the case of stereotactic radiosurgery and
Abstract—The calculation of shielding barrier thicknesses for
radiation therapy facilities according to the NCRP formalism
is based on the use of broad beams (that is, the maximum
possible field sizes). However, in practice, treatment fields used
in radiation therapy are, on average, less than half the
maximum size. Indeed, many contemporary treatment techniques call for reduced field sizes to reduce co-morbidity and
the risk of second cancers. Therefore, published tenth value
layers (TVLs) for shielding materials do not apply to these
very small fields. There is, hence, a need to determine the
TVLs for various beam modalities as a function of field size.
The attenuation of 60Co gamma rays and photons of 4, 6, 10,
15, and 18 MV bremsstrahlung x ray beams by concrete has
been studied using the Monte Carlo technique (MCNP version
4C2) for beams of half-opening angles of 0°, 3°, 6°, 9°, 12°, and
14°. The distance between the x-ray source and the distal
surface of the shielding wall was fixed at 600 cm, a distance
that is typical for modern radiation therapy rooms. The
maximum concrete thickness varied between 76.5 cm and
151.5 cm for 60Co and 18 MV x rays, respectively. Detectors
were placed at 630 cm, 700 cm, and 800 cm from the source.
TVLs have been determined down to the third TVL. Energy
spectra for 4, 6, 10, 15, and 18 MV x rays for 10 ⴛ 10 cm2 and
40 ⴛ 40 cm2 field sizes were used to generate depth dose curves
in water that were compared with experimentally measured
values.
Health Phys. 92(5):456 – 463; 2007
Key words: shielding; accelerators, medical; Monte Carlo;
photons
INTRODUCTION
RADIATION THERAPY rooms are generally constructed of
concrete because of its relatively low cost. To design the
shielding, one determines the thickness of the various
* Department of Physics, University of Massachusetts Lowell,
One University Avenue, Lowell, MA 01854; † Department of Radiation Oncology, Massachusetts General Hospital, Harvard Medical
School, Boston, MA 02114.
For correspondence contact: P. J. Biggs, Department of Radiation
Oncology, Massachusetts General Hospital, Harvard Medical School,
Boston, MA 02114, or email at pjbiggs@partners.org.
(Manuscript accepted 21 November 2006)
0017-9078/07/0
Copyright © 2007 Health Physics Society
456
Tenth value layers ● A. K. JARADAT
stereotactic radiotherapy, both intra- and extra-cranial,
the fields are limited to a few cm in diameter, at most. In
accelerated partial breast irradiation, it was recognized
that, in a select group of patients, the treatment field need
only cover the area of the primary disease and not the
whole breast to achieve similar outcomes (Schwartz et al.
2005). This has reduced the sizes of the fields considerably.
It is now well accepted that Monte Carlo (MC)
methods offer the most powerful tool for modeling
radiation transport for radiotherapy applications (Verhaegen and Seuntjens 2003). Ding (2002a) used the MC
method to simulate radiotherapy photon beams emerging
from an accelerator; using the MC beam code, Hartman
Siantar et al. (2001) has modeled the treatmentindependent accelerator head, resulting in the identification of primary and scattered photon sources and electron
contamination source; Sheikh-Bagheri and Rogers
(2002) used the MC technique to determine the energy
spectra of different Varian linear accelerators.
In radiation shielding applications, where experiments are either very difficult or extremely time consuming, the MC approach has been frequently used to solve
difficult problems. Shobe et al. (1999) measured the
scatter fractions of 6, 10, 18, and 25 MV x rays, while
Nogueira and Biggs (2001) measured the scatter fractions for 4, 6, 10, and 23 MV x rays and also the TVLs
in lead for 4, 6, and 10 MV (Nogueira and Biggs 2002).
Maruyama et al. (1971) evaluated the broad-beam attenuation coefficients for lead, iron, heavy concrete, and
concrete for x-ray beams and compared the results with
experimental measurements. Biggs (1991a) used the MC
program, ITS, to determine the dose and energy spectrum
of radiation at the door for 4 and 10 MV x rays and to
determine the attenuation 60Co gamma ray and photons
of 4, 10, and 18 MV x-ray beams by concrete, steel, and
lead (Biggs 1996). Biggs (1998) also used the MC code
to calculate the shielding thickness required for the roof
of a single-story building. In a similar vein, Kong et al.
(2005) used the MC method to predict skyshine radiation
from electron accelerators and compared the results with
experimental measurements. Also, Casanova et al.
(2004) used the MC method to analyze the shielding
requirements for a MT-25 microtron medical accelerator.
Taylor et al. (1999) have determined the scatter fractions
of dose for a wide range of energies using the MC
methodology. More specifically, Nelson and LaRiviere
(1984) used the EGS MC code to determine the spectra
and transmission properties of 6, 10, and 25 MV bremsstrahlung beams. These transmission data were for incident broad beams (30° opening angle), approximately
equivalent to the beam passing through the primary
AND
P. J. BIGGS
457
collimator in a modern medical linear accelerator, at
azimuthal angles of 0° and about 45°, 90°, and 135°.
The purpose of this paper is to determine the TVLs
in concrete for a range of energies currently used in
radiation therapy as a function of half-opening angle of
the x-ray beam for the first, second and third TVLs.
TVLs for 60Co also were determined because this modality has been well-established in terms of measured data.
MC simulations of photon/electron transport in a
water phantom were also used to obtain depth dose
curves in water. The agreement between the calculation
and the measurement was within the statistical uncertainty of the MC simulations and measured data.
METHODS AND MATERIALS
Monte Carlo code
The MCNP Monte Carlo transport code system,
developed at Los Alamos and distributed by the Radiation Shielding Information Computational Center at Oak
Ridge National Laboratory, was used in this work.
MCNP is a general-purpose Monte Carlo N-particle code
(version 4C2) that can be used for neutron, photon, and
electron, or coupled neutron/photon/electron transport.
For photons, the code takes account of incoherent and
coherent scattering, the possibility of fluorescent emission after photoelectric absorption, and bremsstrahlung
(MCNP 2003).
In this study, because of the small number of
photoneutrons produced in the interaction of photons
with concrete that produce capture gamma rays, only
coupled photons and electrons were studied with no
neutrons included. In addition, the version of MCNP
used in this study did not have the gamma-neutron
cross-sections as a secondary source and therefore the
process could not be simulated (Casanova et al. 2004).
The results in this paper are based on the photon flux,
determined using the f4 tally. This permits a direct
comparison with the flux striking the barrier and the ratio
defines the transmission factor. It can be argued that the
spectrum of photons exiting the barrier is considerably
different (lower average energy) from the spectrum of
photons entering the barrier. However, this is in the spirit
of the definition of the transmission factor (Johns and
Cunningham 1983). If one were to use an f8 tally, energy
deposited or dose, one still has to face the problem of a
different energy spectrum before and after the barrier.
Additionally, there is probably not electronic equilibrium
for the spectrum of photons exiting the barrier, and one
has to determine the dose of the incoming beam, which
could be arbitrary since it can be defined at different
depths in a phantom.
458
Health Physics
The electron energy cut-off was set to 50 keV and
the photon energy cut-off to the default of 1 keV. These
cut-offs could have been set higher in the barrier, for
example, without affecting the final result. The runs
were, consequentially, longer.
Barrier material
Concrete was used for the barrier shielding material
with an assumed density of 2.35 g cm⫺3. The weight
composition of concrete used in this calculation was
taken to be hydrogen, 0.55%; oxygen, 49.57%; sodium,
1.7%; magnesium, 0.26%; aluminum, 4.55%; silicon,
31.36%; sulfur, 0.13%; potassium, 1.9%; calcium,
8.26%; and iron, 1.23% (NCRP 1977).
Tenth value layer (TVL)
In radiation shielding work, it is customary to use the
TVL of a material, which is defined as the thickness of the
material needed to reduce the intensity of radiation to one
tenth of the initial intensity of the beam at the point of
measurement. Because of the build-up factor that occurs for
broad beams, the first TVL is always greater than the
second and third TVLs. By measuring or calculating the
attenuation as a function of absorber thickness, one can
determine the required thickness of shielding material for
the primary barrier in any radiotherapy room.
Bremsstrahlung beam
The bremsstrahlung spectra for 4, 6, 10, 15, and 18
MV were generated for 10 ⫻ 10 cm2 for Varian machines
by Sheikh-Bagheri and Rogers (2002) using the MC BEAM
code. The most probable energies for these beams were 1.0,
0.75, 1.125, 1.5, and 1.75 MeV, and the average energies
were 1.52, 1.82, 3.19, 3.88, and 4.88 MeV, respectively.
Geometry
Transmission measurements. The geometry used in
this MC simulation consisted of a point source x-ray beam
incident on the absorbing material within different coneshaped angles with circular cross sections. Whereas the
maximum field size for conventional linear accelerators is
square,‡ it is much simpler to generate a field of circular
cross-sections in a MC simulation. Table 1 shows a list of
the equivalent square fields for each different cone angle
field at 1 m from the x-ray target. The layout of this
experiment is shown in Fig. 1. The transmitted radiation
was detected with a spherical air-filled detector with a 3-cm
radius on the beam central axis at distances of 630 cm, 700
cm, and 800 cm from the source and therefore at different
distances from the barrier, namely 30 cm, 100 cm, and 200
‡
Note that this is not strictly correct for most machines since the
field is also limited by a circular, fixed primary collimator that projects
a field size of 50 cm diameter at isocenter.
May 2007, Volume 92, Number 5
Table 1. Equivalent square area at 1 m from the x-ray target.
Angle (°)
Area (cm2)
0
3
6
9
12
14
0
9.3 ⫻ 9.3
18.6 ⫻ 18.6
28.1 ⫻ 28.1
37.7 ⫻ 37.7
44.2 ⫻ 44.2
Fig. 1. Source, concrete barrier and the three positions of the
detector (a, b, c) behind the barrier. Point a is 30 cm behind the
barrier, point b 100 cm, and point c 200 cm.
cm, respectively. This was done to evaluate the variation of
TVL with distance between the detector and the barrier. Air
was assumed as the medium for all regions other than the
barrier.
Depth dose measurements. Depth doses were generated for these beams to ensure that the applied MC
methodology was correct. The geometry used to generate
the percent depth dose curves consisted of a point source,
cylindrical chamber of radius of 1 cm and thickness of 0.2
cm, and a phantom water tank of dimensions 1 ⫻ 1 ⫻ 1 m3.
The distance between the source and the water surface was
1 m (i.e., the water surface was at the isocenter), and the
chamber had its flat surface parallel to the water surface.
The percent depth dose curves were simulated for two field
sizes: 10 ⫻ 10 cm2 and 40 ⫻ 40 cm2.
Data collection and normalization. The transmitted
radiation was sampled by measuring the number of photons
per photon emitted at the target for each cone angle and
energy. Results were normalized to the case with no
shielding material present. The transmission was measured
down to ⬍10⫺3 to ensure an adequate measurement of the
third TVL. The errors on the transmission points ranged
from ⬍⬍1% for shallow thicknesses and small cone opening
Tenth value layers ● A. K. JARADAT
AND
P. J. BIGGS
459
angles, to ⬃5% for transmission factors of ⬍10⫺3 and large
cone opening angles. These numbers were essentially independent of beam energy. The overall average error was
2.2% with a standard deviation of 1.0%.
RESULTS
Figs. 2 and 3 are plots of the percent depth dose
derived from the spectra used for simulation vs. experimental measurements for a field size of 40 ⫻ 40 cm2 at
6 MV and 10 ⫻ 10 cm2 at 18 MV, respectively. In each
case the data points correspond to the MC calculation,
and the line corresponds to measured data.
The transmission data consist of 18 plots for various
detector distances from the barrier and different energies
(3 distances ⫻ 6 energies); each plot contains one
contour for each of the six cone angles. It would be
impractical to show all plots. Instead, a sample of these
plots is shown to illustrate the type of curves generated.
For each plot, the transmission is plotted on a log scale
against barrier thickness, and the lines simply bridge the
data points to aid the eye.
Figs. 4 –7 are families of transmission curves through
concrete for cone angles between 0° and 14° for 60Co and 18
MV x rays at 30 cm and 200 cm detector-barrier distances,
respectively. Note that the curves in Figs. 4 and 5 for 60Co
gamma rays, for angles other than 0°, are not purely
exponential, but have a convex shape or shoulder at shallow
depths, corresponding to the build-up factor (Chilton et al.
1984). Build-up is also visible at 18 MV for the 30 cm data.
The effect almost vanishes at 18 MV for 200 cm. Analysis
Fig. 3. Percent depth dose curve 18 MV 10 ⫻ 10 cm2 field size (ⴱ)
MCNP; (——) experiment.
Fig. 4. Transmission curves for 60Co gamma rays in concrete at a
30 cm detector barrier distance. (⫹) 0°; (‚) 3°; (䡺) 6°; (䉬) 9°; (ⴱ)
12°; (Œ) 14°.
Fig. 2. Percent depth dose curve for 6 MV 40 ⫻ 40 cm2 field size
(ⴱ) MCNP; (——) experiment.
of these data show that the build-up is greater for 60Co than
for 18 MV, but, to the eye, the effect is small. Also
noticeable is that, for both energies, the transmission curve
for 0° is widely separated from the other curves for the 30
cm data but not the 200 cm data.
The variation of the first and third TVLs, respectively,
as a function of energy, is shown in Figs. 8 and 9,
respectively, for all cone angles at 30 cm, compared with
the first and subsequent TVLs of NCRP Report No. 51
(NCRP 1977) (heavy dashed line) and NCRP Report No.
151 (NCRP 2005) (heavy dotted line). The first TVL for all
angles at 200 cm, again as a function of energy, are
compared with the first TVL of NCRP Report No. 51
460
Health Physics
Fig. 5. Transmission curves for 60Co gamma rays in concrete at a
200 cm detector barrier distance. (⫹) 0°; (‚) 3°; (䡺) 6°; (䉬) 9°;
(ⴱ) 12°; (Œ) 14°.
May 2007, Volume 92, Number 5
Fig. 7. Transmission curves for 18 MV x rays in concrete at a 200
cm detector to barrier distance. (⫹) 0°; (‚) 3°; (䡺) 6°; (䉬) 9°; (ⴱ)
12°; (Œ) 14°.
Fig. 6. Transmission curves for 18 MV x rays in concrete at a 30
cm detector to barrier distance. (⫹) 0°; (‚) 3°; (䡺) 6°; (䉬) 9°; (ⴱ)
12°; (Œ) 14°.
(NCRP 1977) and NCRP Report No. 151 (NCRP 2005) in
Fig. 10. The NCRP values refer to broad beam transmission
values and are assumed to represent the largest field size.
Tables 2– 4 provide the first, second, and third TVLs
for all energy and opening angle combinations. Using the
uncertainties on the individual measurements given earlier,
an error analysis was performed on the TVL values. Data
were analyzed for all energies at 0°, 6°, and 14°. For the
first, second, and third TVLs, the error is negligible (⬃0.1
cm) for the 0° cone angle, and it increases with cone angle
and from the first to the third TVLs. Averaged over all
distances (30 cm, 100 cm, and 200 cm), the error increased
from 0.1 cm to 0.7 cm at 6° and to 1.0 cm at 14°. The error,
averaged over all energies and distances, increased from 0.3
cm for the first TVL to 0.6 cm for the second TVL and to
0.8 cm for the third TVL. A histogram of all the points
analyzed gives a most probable error of 0.2 cm and an
Fig. 8. First TVLs in concrete as a function of energy for a 30 cm
barrier to detector distance. (⫹) 0°; (‚) 3°; (䡺) 6°; (䉬) 9°; (ⴱ) 12°;
(Œ) 14°; (– – –) NCRP 51; (- - -) NCRP 151.
average of 0.5 ⫾ 0.5 cm. If one were to fit the data with a
model, one could likely get smaller errors, but these would
then be model dependent. The analysis given above therefore represents a safe upper limit.
DISCUSSION
The good agreement between the MCNP-generated
percent depth dose curves and the measured data is demonstrated in Figs. 2 and 3. Note that a full MC simulation of
the treatment head was not used in this work since that
would be unnecessary, so some discrepancies are to be
expected. For example, although photons scattered by the
jaws would be included in the simulation, in this study we
Tenth value layers ● A. K. JARADAT
Fig. 9. Third TVLs in concrete as a function of energy for a 30 cm
barrier to detector distance. (⫹) 0°; (‚) 3°; (䡺) 6°; (䉬) 9°; (ⴱ) 12°;
(Œ) 14°; (– – –) NCRP 51; (- - -) NCRP 151.
Fig. 10. TVLs in concrete as a function of energy for a 200 cm
barrier to detector distance. (⫹) 0°; (‚) 3°; (䡺) 6°; (䉬) 9°; (ⴱ) 12°;
(Œ) 14°; (– – –) NCRP 51; (- - -) NCRP 151.
only used the photon spectra emanating from a point,
namely the target. There is a discrepancy between the two
sets of data in the build-up region for 18 MV, as expected,
since electron contamination is a well-known effect at this
energy. The effect is much less at lower energies, which
explains the better agreement at 6 MV, despite the larger
field size that will increase the effect of contamination. This
MC simulation does not model electron contamination from
the treatment head, only that generated in the intervening
air. There is good agreement beyond the depth of the
maximum dose. The discrepancy in the build-up region was
AND
P. J. BIGGS
461
investigated by Ding (2002b), who was unable to solve the
problem. He added a lead filter to reject electrons in the
build-up region, but was unable to reconcile the measured
data with his MC data.
The effect of build-up, for angles of 9° and higher, can
be seen in the transmission plots at 30 cm detector-barrier
distance. However, at the 200 cm detector-barrier distance,
the build-up effect is considerably less, and largely absent at
18 MV because the scattering source in the concrete is
farther away from the detector and therefore will make a
lower contribution. Note that for 0° the curves for each
energy are independent of distance, as expected, since no
in-scatter is involved. This also serves as a secondary check
on the applied MC methodology. Further, the curves are not
purely exponential, having a larger slope (lower TVL) at the
beginning compared with greater depths. These curves are
akin to those seen in diagnostic radiology and the first part
of the curve represents the attenuation of the low energy
component of the beam.
A general pattern for the TVLs listed in Tables 2, 3,
and 4, for all detector-barrier distances, is that they
increase with increasing energy and cone angle. The
dependence on cone angle is because of the increased
scatter contribution, and it shows how the build-up
occurs, i.e., no build-up for small field size, but increasing with increasing angle.
The first TVL in Fig. 8 at 30 cm detector-barrier
distance is higher than those given in NCRP Report Nos. 51
and 151 for cone angles greater than 6°, although there is
better agreement with the data from NCRP Report No. 151
at 4 and 6 MV. However, the MC data are in good
agreement with the NCRP Report Nos. 51 and 151 data for
the third TVL, as shown in Fig. 9. Interestingly, the first
TVL data agree well with the NCRP data at a barrierdetector distance of 200 cm (Fig. 10), but less so with the
data from NCRP Report No. 151 at 4 and 6 MV. It is
unclear why the first TVL calculated by MC is greater than
the published values. However, as the footnote in NCRP
Report No. 51 states, “curves are empirically drawn through
data points” (NCRP 1977),” and it is clear that no data are
shown between 2 and 35 MV of the subsequent TVL curve.
However, NCRP Report No. 151 provides more recent data
for these specific machine energies. On the other hand, the
200 cm data show that the measurements might have been
performed at a distance much greater than 30 cm. This is
highlighted in the case of 60Co where the difference between
the MC and NCRP data is particularly significant for the
first TVL, whereas for the third TVL there is excellent
agreement between the MC data and that from NCRP
Report No. 151. Further analysis of the 60Co data shows that
all the second and third TVLs have about the same values,
consistent with the NCRP values, with a slow increase from
⬃17.5 cm at 0° to ⬃21 cm at 14°. In contrast, the first TVLs
462
Health Physics
May 2007, Volume 92, Number 5
Table 2. First TVL at 0.3, 1.0, and 2.0 m detector-barrier distance for different cone angles.
Cone angle (°)
Energy (MV)
Distance (m)
1.25
1.25
1.25
4.00
4.00
4.00
6.00
6.00
6.00
10.00
10.00
10.00
15.00
15.00
15.00
18.00
18.00
18.00
0.3
1.0
2.0
0.3
1.0
2.0
0.3
1.0
2.0
0.3
1.0
2.0
0.3
1.0
2.0
0.3
1.0
2.0
0°
3°
6°
9°
12°
14°
17.5
17.2
17.0
18.0
18.0
18.0
20.0
19.5
19.5
25.5
25.7
25.5
28.0
27.5
28.0
31.0
29.4
30.0
28.0
21.0
18.5
29.5
22.5
19.5
32.0
24.5
21.5
40.5
32.0
28.5
44.0
34.5
31.5
48.0
38.2
34.5
32.5
26.2
22.5
33.5
27.6
23.5
37.0
30.3
26.0
45.5
38.7
34.0
50.0
41.7
37.5
54.5
46.2
41.0
33.5
29.8
26.0
35.5
31.5
27.5
38.5
34.3
30.0
47.5
43.0
38.5
52.0
47.0
42.0
56.5
50.6
48.0
34.5
31.6
29.0
36.5
33.6
30.5
40.0
36.0
33.0
50.5
46.0
42.0
53.0
48.7
45.0
58.0
53.6
50.0
34.5
32.5
29.5
36.5
33.8
31.0
40.0
37.5
34.5
50.5
47.0
43.5
53.0
49.5
46.0
58.5
54.6
51.0
Table 3. Second TVL at 0.3, 1.0, and 2.0 m detector-barrier distance for different cone angles.
Cone angle (°)
Energy (MV)
Distance (m)
1.25
1.25
1.25
4.00
4.00
4.00
6.00
6.00
6.00
10.00
10.00
10.00
15.00
15.00
15.00
18.00
18.00
18.00
0.3
1.0
2.0
0.3
1.0
2.0
0.3
1.0
2.0
0.3
1.0
2.0
0.3
1.0
2.0
0.3
1.0
2.0
0°
3°
6°
9°
12°
14°
17.0
17.2
17.5
20.5
20.4
20.0
24.5
24.8
24.5
32.0
31.8
31.5
34.0
34.0
33.5
36.0
38.2
36.5
20.5
20.6
19.0
25.5
24.0
22.5
31.0
28.8
27.0
39.5
37.0
34.5
40.5
39.5
36.0
42.5
42.2
40.0
21.5
21.6
20.0
26.0
26.5
25.5
31.5
30.7
29.5
41.0
39.6
38.0
42.0
42.0
40.0
43.5
43.8
42.0
22.0
21.8
21.5
26.0
26.5
25.5
31.5
31.4
30.0
40.0
39.5
39.5
43.0
42.0
41.0
43.5
44.4
42.5
22.0
21.9
21.0
26.5
26.6
26.0
31.5
32.5
30.5
39.0
40.0
38.5
43.0
42.8
41.0
44.0
44.0
42.5
22.0
22.0
21.5
26.5
26.6
27.0
31.5
32.5
31.0
40.0
40.0
38.5
43.0
43.0
41.0
44.0
44.0
43.0
Table 4. Third TVL at 0.3, 1.0, and 2.0 m detector-barrier distance for different cone angles.
Cone angle (°)
Energy (MV)
Distance (m)
1.25
1.25
1.25
4.00
4.00
4.00
6.00
6.00
6.00
10.00
10.00
10.00
15.00
15.00
15.00
18.00
18.00
18.00
0.3
1.0
2.0
0.3
1.0
2.0
0.3
1.0
2.0
0.3
1.0
2.0
0.3
1.0
2.0
0.3
1.0
2.0
0°
3°
6°
9°
12°
14°
18.0
17.4
17.5
23.5
23.0
23.0
28.0
28.0
28.0
36.0
35.5
35.5
38.0
37.5
37.5
40.0
39.4
40.0
19.0
19.6
19.0
26.0
25.2
25.0
31.5
31.5
30.5
38.0
38.5
39.0
41.5
41.8
41.5
45.5
43.2
42.5
19.5
20.8
20.5
27.0
26.2
26.0
33.5
33.0
32.5
39.0
40.2
40.5
42.0
42.0
41.5
46.0
45.0
45.5
20.5
20.6
21.5
27.5
26.5
28.0
34.0
33.0
34.5
40.5
40.5
42.0
42.5
42.3
42.0
46.0
45.2
46.0
20.5
20.5
21.0
27.5
27.3
28.0
34.5
33.5
34.5
40.0
40.5
41.0
44.0
42.5
43.0
48.0
47.4
46.5
21.0
20.5
20.5
27.5
28.0
28.0
34.5
33.5
36.5
40.0
41.3
40.5
45.0
44.0
43.0
49.0
47.4
47.0
Tenth value layers ● A. K. JARADAT
have a much larger TVL for large angles, irrespective of
distance.
CONCLUSION
The MCNP program has been used to study the
TVLs for energies between 60Co gamma rays and 18 MV
bremsstrahlung incident on concrete at various cone
angles up to 14°. The data at large angles agree with
NCRP Report No. 151. For smaller angles, the TVL
values are significantly less. The results, for cone angles
less than the maximum, show that the TVLs are lower
than for broad beams. Since average field sizes in
radiation therapy are considerably less than the 40 ⫻ 40
cm2 that the machines allow, thinner shielding walls than
calculated using the standard NCRP values could be
used. For machines that are dedicated to specific procedures, such as stereotactic radiotherapy, this change
could be quite significant in view of the small field sizes
used in that procedure. Even a machine dedicated to
breast cancer, as in our institution, would use considerably smaller field sizes. As an aside, field size reduction
would also reduce the scatter dose considerably since it
varies as the area of the field (Shobe et al. 1999).
REFERENCES
Biggs PJ. Calculation of shielding door thicknesses for radiation therapy facilities using the ITS Monte Carlo program.
Health Phys 61:465– 472; 1991a.
Biggs PJ. Analysis of field and custom block sizes used in
radiation therapy—implications for multileaf collimator
design. Med Dosim 16:169 –172; 1991b.
Biggs PJ. Obliquity factors for Co-60, and 4, 10, and 18 MV x
rays for concrete, steel, and lead and angles of incidence
between 0° and 70°. Health Phys 70:527–536; 1996.
Biggs PJ. Determination of shielding for an unoccupied roof of
a radiation therapy facility to account for adjacent buildings—modification of a computer program. Health Phys
75:431– 434; 1998.
Casanova AO, Lopez N, Gelen A, Guevaro MV, Diaz O,
Cimino L, D’Alessandro K, Melo JC. Shielding analysis of
the microtron MT-25 using the MCNP 4C code and NCRP
Report No. 51. Radiat Protect Dosim 109:189 –195; 2004.
Chilton, AB, Shultis JK, Faw RE. Principles of radiation
shielding. Englewood Cliffs, NJ: Prentice-Hall, Inc.; 1984.
Ding GX. Energy spectra, angular spread, fluence profiles and
dose distributions of 6 and 18 MV photon beams: results of
Monte Carlo simulations for a Varian 2100EX accelerator.
Phys Med Biol 47:1025–1046; 2002a.
Ding GX. Dose discrepancies between Monte Carlo calculations and measurements in the buildup region for a high
energy photon beam. Med Phys 29:2459 –2463; 2002b.
AND
P. J. BIGGS
463
Hartman Siantar CL, Walling RS, Daly TP, Faddegon B, Albright
N, Bergstrom P, Bielajew AF, Chuang C, Garrett D, House
RK, Knapp D, Wieczorek DJ, Verhey LJ. Description and
dosimetric verification of the PEREGRINE Monte Carlo dose
calculation system for photon beams incident on a water
phantom. Med Phys 28:1322–1337; 2001.
Johns HE, Cunningham JR. Physics of radiology. Springfield,
IL: Charles C. Thomas; 1983.
Kong C, Li Q, Chen H, Du T, Cheng C, Tang C, Li Z, Zhang
H, Pei Z, Ming S. Monte Carlo method for calculating the
radiation skyshine produced by electron accelerators. Nucl
Instr Meth Phys Res B 234:269 –274; 2005.
Maruyama T, Kumamoto Y, Kato Y, Hashizume T, Yamamoto
M. Attenuation of 4 –32 MV x-rays in ordinary concrete, heavy
concrete, iron, and lead. Health Phys 20:277–284; 1971.
MCNP. A general Monte Carlo N-particle transport code,
version 4C2. Los Alamos: Los Alamos National Laboratory; Report LA-UR-03-1987; 2003.
National Council on Radiation Protection and Measurements.
Structural shielding design and evaluation for medical use
of x-rays and gamma rays of energies up to 10 MV.
Bethesda, MD: NCRP; NCRP Report No. 49; 1976.
National Council on Radiation Protection and Measurements.
Structural shielding design and evaluation for medical use
of x-rays and gamma rays of energies up to 10 MV.
Bethesda, MD: NCRP; NCRP Report No. 51; 1977.
National Council on Radiation Protection and Measurements.
Neutron contamination from medical electron accelerators.
Bethesda, MD: NCRP; NCRP Report No. 79; 1984.
National Council on Radiation Protection and Measurements.
Structural shielding design and evaluation for megavoltage
x- and gamma-ray radiotherapy facilities. Bethesda, MD:
NCRP; NCRP Report No. 151; 2005.
Nelson WR, LaRiviere PD. Primary and leakage radiation calculations at 6, 10 and 25 MV. Health Phys 47:811– 818; 1984.
Nogueira IP, Biggs PJ. Measurement of scatter factors for 4, 6,
10 and 23 MV x rays at scattering angles between 30° and
135°. Health Phys 81:330 –340; 2001.
Nogueira IP, Biggs PJ. Measurement of TVLs in lead for 4, 6
and 10 MV bremsstrahlung x-ray beams at scattering angles
between 30° and 135°. Health Phys 83:255–260; 2002.
Schwartz GF, Veronesi U, Clough K, Dixon JM, Fentiman IS,
Heywang-Köbrunner SH, Holland R, Hughes KS, Mansel
RE, Margolese EB, Olivotto IA, Palazzo JP, Solin LJ.
Consensus conference on breast conservation. J Am Coll
Surgeons 203:198 –207; 2005.
Sheikh-Bagheri D, Rogers DWO. Monte Carlo calculation of
nine megavoltage photon beam spectra using the BEAM
code. Med Phys 29:391– 402; 2002.
Shobe J, Rodgers JE, Taylor PL. Scattered fractions from 6, 10,
18 and 25 MV linear accelerator x-rays in radiotherapy
facilities. Health Phys 76:27–35; 1999.
Taylor PL, Rogers JE, Shobe J. Scatter fractions from linear
accelerators from 6 to 24 MV. Med Phys 26:1442–1446; 1999.
Verhaegen F, Seuntjens J. Monte Carlo modeling of external
radiotherapy photon beams. Phys Med Biol 48:R107–164;
2003.
f f